Download Personal Surveyor 200 User Handbook PDF

TitlePersonal Surveyor 200 User Handbook
LanguageEnglish
File Size6.0 MB
Total Pages96
Table of Contents
                            TABLE OF CONTENTS
	COPYRIGHT
	LIABILITY
	MODIFICATION NOTICES
	SOFTWARE
	DISPOSAL ADVICE
	SAFETY
	AREAS OF USE
	SPECIAL CONDITIONS OF USE
	STORAGE, HANDLING AND TRANSIT
	WARRANTY
	DECLARATION OF CONFORMITY
	REVISION RECORD
	INTRODUCTION
	1.1  GENERAL DESCRIPTION
	1.2  FEATURES
	1.3  DATA LOGGING
	1.3.1  Archiving Stored Readings
	1.4  FILTERS
	1.5  CONSTRUCTION
	1.6  IDENTIFICATION LABEL
	1.7  CERTIFICATION
	1.7.1  Certification Marks
	1.7.2  Performance
	OPERATION
	2.1  OPERATING PROCEDURE
	2.2  SWITCH THE INSTRUMENT ON
	2.2.1  Instrument Identification
	2.2.2  Battery Status
	2.2.3  User Name / Number Only (Option)
	2.2.4  'Man Down' (Motion Sensor) Alarm Option
	2.2.5  Date and Time
	2.2.6  Bump Due Date (Option)
	2.2.7  Calibration Due Date
	2.2.8  Service Due Date (Option)
	2.2.9  Select Calibration Gas (Option)
	2.2.10  Sensor Confirmation Check
	2.2.11  Memory Fault
	2.2.12  Normal Operating Display
	2.3  SWITCH THE DISPLAY BACKLIGHT ON / OFF
	2.4  MANUAL DATALOG
	2.5  VIEW MAXIMUM AND MINIMUM RECORDED VALUES SINCE SWITCH ON
	2.6  ALARMS RESET OR ACKNOWLEDGE
	2.6.1  Confidence Signal
	2.7  REMOTE SAMPLING (with pump option)
	2.7.1  Pump Operation:
	2.8  SELF TEST
	2.9  SWITCH THE INSTRUMENT OFF
	MANUAL BUMP TEST OPTIONS
	3.1  BACKGROUND
	3.2  MANUAL BUMP OPTIONS
	3.3  INITIATING A MANUAL BUMP TEST
	3.4  APPLYING TEST GAS
	3.5 QUICK / FULL BUMP TEST
	3.5.1  Quick Bump Test
	3.5.2 Full Bump Test
	3.6  ALARM CONFIRMATION
	3.7  BUMP TEST RESULT
	ALARMS
	4.1  GAS ALARMS
	4.1.1 Flammable LEL Alarm Limit
	4.1.2  Over-Range Flammable Gas Alarm Function
	4.1.3  Oxygen (O2) Alarm Limits
	4.1.4  Toxic Alarm Limits
	4.2  ACKNOWLEDGE GAS ALARMS
	4.3  MUTE ALARMS
	4.4  HIGH FLAMMABLE GAS OVER-RANGE ALARM
	4.5  FAULT ALARMS
	4.5.1  Low Battery
	4.5.2  Zero Fault
	4.5.3  Sensor Fault
	4.5.4  Low Flow - (Pumped instruments only)
	4.5.5  Flow Fault (Pumped Instruments Only)
	4.5.6  Calibration Required
	4.5.7  Calibration Expired
	4.6  'MAN DOWN' (MOTION SENSOR) ALARM
	OPERATOR MAINTENANCE
	5.1  CLEANING
	5.2  REPLACE INSTRUMENT FILTERS
	5.2.1  Replace Sensor Hydrophobic Filter
	5.2.2  Replace Sample Inlet (Dust) Filter
	5.2.3  In-line Hydrophobic Filter (Accessory)
	RECHARGE BATTERY
	6.1  RECHARGE INSTRUMENT BATTERY
	6.1.1  Recharge Instrument using the Charging / Comms Clip
	6.1.2  Recharge Instrument using the 5-way Charger
	6.1.3  Recharge Instrument using the 12V / 24V In-Vehicle Charging Adaptor
	CALIBRATION
	7.1  GENERAL DESCRIPTION
	7.2  CALIBRATION VALIDITY
	ACCESSORIES
	ADDITIONAL INFORMATION
	9.1  TRAINING
	9.2  WORLD WIDE WEB
	TYPICAL OPERATING PARAMETERS
	PHYSICAL PROPERTIES
	ENVIRONMENT
	TYPICAL FLOW RATE INFORMATION
	WARM-UP / STABILIZATION TIME
	ALARMS
	DISPLAY
	POWER SOURCE
	CONSTRUCTION
	IP RATING
	RESPONSE TIMES (T90)
	INDEX
INDEX
	Symbols
		5-way Charger
	A
		ACCESSORIES
		ACKNOWLEDGE
		ACKNOWLEDGE GAS ALARMS
		ADDITIONAL  INFORMATION
		ALARM CONFIRMATION
		ALARMS
		ALARMS, FAULT
		ALARMS, GAS
		ALARMS, MUTE
		ALARMS RESET OR ACKNOWLEDGE
		APPLYING TEST GAS
		AREAS OF USE
		Automatic Calibration
	B
		BACKLIGHT, DISPLAY
		BAT
		BATTERY, INSTRUMENT
		Battery, Low
		BATTERY, RECHARGE
		Battery Status
		BUMP OPTIONS
		BUMP TEST OPTIONS
		BUMP TEST RESULT
	C
		CALIBRATION
		Calibration, Automatic
		Calibration Due Date
		Calibration Expired
		Calibration, Field
		Calibration Gas
		Calibration Required
		CALIBRATION VALIDITY
		CERTIFICATION
		Certification Marks
		Charger, 5-way
		CLEANING
		CONFIDENCE BEEP
		Confidence Signal
		CONFIRMATION, ALARM
		Conformity, Declaration
		CONSTRUCTION
		COPYRIGHT
		CSA
	D
		DATA LOGGING
		Date
		Declaration of Conformity
		DESCRIPTION,  GENERAL
		DISPLAY
		DISPLAY BACKLIGHT ON / OFF
		Display, Operating
		DISPOSAL ADVICE
		DOWN, MAN
		Down, Man
	E
		EN50054
		ENVIRONMENT
		Expired, Calibration
	F
		FAULT ALARMS
		Fault, Flow
		Fault, Memory
		Fault, Sensor
		Fault, Zero
		FEATURES
		Field Calibration
		Filter, Hydrophobic
		FILTERS
		FILTERS, INSTRUMENT
		Flammable LEL Alarm Limit
		Flow Fault
		Flow, Low
		FLOW RATE  INFORMATION
		FULL BUMP TEST
	G
		GAS ALARMS
		Gas, Calibration
		GAS OVER-RANGE ALARM
		GAS, TEST
		GENERAL DESCRIPTION
	H
		HANDLING
		HIGH FLAMMABLE GAS OVER-RANGE ALARM
		Hydrophobic Filter
	I
		Identification, Instrument
		IDENTIFICATION LABEL
		In-line Hydrophobic Filter (Accessory)
		In-Vehicle Charging Adaptor
		INFORMATION,  ADDITIONAL
		INITIATING A MANUAL BUMP TEST
		Inlet (Dust) Filter
		INSTRUMENT BATTERY
		INSTRUMENT FILTERS
		Instrument Identification
		INSTRUMENT OFF
		INSTRUMENT ON
		INTRODUCTION
		IP RATING
		IP65
	L
		LABEL, IDENTIFICATION
		LEL Alarm Limit
		LIABILITY
		LOGGING, DATA
		Low Battery
		Low Flow
		LTEL
	M
		MAINTENANCE,  OPERATOR
		MAN DOWN
		Man Down
		MANUAL BUMP  OPTIONS
		MANUAL BUMP TEST
		MANUAL BUMP TEST OPTIONS
		Marks, Certification
		MAXIMUM
		Memory Fault
		MINIMUM
		MODIFICATION  NOTICES
		MOTION SENSOR
		Motion Sensor
		MUTE ALARMS
	N
		Name, User
		Normal Operating Display
	O
		OFF, INSTRUMENT
		ON, INSTRUMENT
		Operating Display
		OPERATING  PARAMETERS
		OPERATING  PROCEDURE
		OPERATION
		OPERATOR  MAINTENANCE
		Option, Pump
		OPTIONS, BUMP
		OVER-RANGE ALARM
		Over-Range Flammable Gas Alarm Function
		Oxygen (O2) Alarm Limits
	P
		PARAMETERS,  OPERATING
		Perforformance
		PHYSICAL PROPERTIES
		POWER SOURCE
		Power Source
		PROCEDURE,  OPERATING
		PROPERTIES, PHYSICAL
		Pump Option
	Q
		QUICK / FULL BUMP TEST
		QUICK BUMP TEST
	R
		RATE, FLOW
		RATING, IP
		RECHARGE BATTERY
		RECHARGE  INSTRUMENT BATTERY
		Recharge Instrument using the 12V / 24V
		Recharge Instrument using the 5-way Charger
		REMOTE SAMPLING
		REPLACE INSTRUMENT FILTERS
		Replace Sample Inlet (Dust) Filter
		Replace Sensor  Hydrophobic Filter
		Required, Calibration
		RESET, ALARMS
		Resolution
		RESPONSE TIMES
		RESULT, TEST
	S
		SAFETY
		SAFETY, ADDITIONAL DETAILS - CSA ONLY
		Sample Inlet (Dust) Filter
		SAMPLING, REMOTE
		Select Calibration Gas
		SELF TEST
		Sensor Confirmation Check
		Sensor Fault
		Sensor Hydrophobic Filter
		SENSOR, MOTION
		Sensor, Motion
		Signal, Confidence
		SOFTWARE
		SOURCE, POWER
		Source, Power
		STABILISATION TIME
		Status, Battery
		STEL
		STORAGE
	T
		TEST, BUMP
		TEST GAS
		TEST RESULT
		TEST, SELF
		Time
		TIME, STABILISATION
		TIMES, RESPONSE
		Toxic Alarm Limits
		TRAINING
		TRANSIT
		TWA
		TYPICAL FLOW RATE INFORMATION
		TYPICAL OPERATING PARAMETERS
	U
		User Name / Number Only (Option)
	V
		VALIDITY, CALIBRATION
		Vehicle Charging Adaptor
	W
		warm-up
		WARM-UP / STABILISATION TIME
		WORLD WIDE WEB
	Z
		Zero Fault
                        
Document Text Contents
Page 2

Issue 6

18/07/2013

Part Number: 64171

GMI welcomes comments on all our publications.
Your comments can be of great value in helping us to
improve our customer publications. Please send any
comments that you have to our Sales Department at
GMI. Contact details are provided inside the back cover
of this handbook.

Instrument Service / Repair contact details are also
provided inside the back cover of this handbook.

Copyright © Gas Measurement Instruments Ltd 2012

Page 48

3-4

USER HANDBOOK

3.5 QUICK / FULL BUMP TEST

This stage of the bump test is dependent on whether QUICK
or FULL bump is configured.

3.5.1 Quick Bump Test

As the alarm threshold for each range is exceeded, the
audible / visual / vibration alarms will activate and a
symbol will appear, otherwise a symbol will be
displayed.

3.5.2 Full Bump Test

After a short period of time, the gas readings are checked
against configurable limits. The audible / visual / vibration
alarms will activate and a symbol will appear if the
readings are within these limits, otherwise a symbol
will be displayed.

3.6 ALARM CONFIRMATION

The user is then prompted to confirm if the audible and
visual alarms were activated, as shown in Fig. 3-4

Fig. 3-4 Confirm Alarms (4-gas model)